Frequently Asked Questions


Procurement and quality questions, answered the way a supplier questionnaire asks them.

Common Questions

If a question below is not covered, it can be answered directly — see the contact page.

  • Do you offer CMM calibration near me if my plant is outside Detroit?

    Coverage runs through Detroit, Toledo, Flint, Jackson, Saginaw, Battle Creek and Grand Rapids, plus the surrounding towns in each area. Machines outside that range are typically handled ship-in; on-site service is confirmed by distance. Routing for a specific address is confirmed through a quote request.

  • How often should a stamping-plant CMM be calibrated?

    Interval is set by the operator's own quality procedure and the machine's duty cycle, typically annually for machines under IATF 16949 monitoring. Heavier use, coolant exposure or a recent crash can justify a shorter interval.

  • What does ISO 10360-2 actually test on a stamping-floor CMM?

    Five calibrated test lengths are measured three times each in seven positions, 105 measurements total, verifying length measurement error E0 and E150 against the machine's stated MPE. Probing error is tested separately under ISO 10360-5.

  • Can calibration be done on site, and how long is the machine down?

    On-site service is available for most bridge and gantry CMMs. Time on site depends on machine size and whether adjustment is needed, and is confirmed on the quote.

  • What temperature does the room need to be for testing?

    ISO 1 sets 20°C as the dimensional reference temperature. Acceptance testing follows the environment specified by the machine's manufacturer; reverification may use conditions the user specifies, which is why the recorded conditions on the certificate matter.

  • What happens if the CMM fails as-found testing?

    As-found results are recorded before adjustment. Notification of an out-of-tolerance finding is sent before correction, with data sufficient to support an impact assessment on parts accepted since the last passing certificate.

  • Is the certificate accredited, and how do I check the scope?

    Certificates are issued under the performing laboratory's ISO/IEC 17025:2017 scope of accreditation. The accrediting body's public directory can be checked against the specific measurement, range and uncertainty before the work is ordered.

  • What is a CMC, and why does it matter to my tolerances?

    CMC, calibration and measurement capability, is the smallest uncertainty a laboratory can claim within its scope. If the CMC does not comfortably support the tolerance being inspected, a 4:1 test uncertainty ratio may not hold, and that affects the conformance decision.

  • Will this satisfy my IATF 16949 auditor?

    IATF 16949 clause 7.1.5.3 requires evidence that an external laboratory is accredited to ISO/IEC 17025 or otherwise acceptable to the customer. Certificates and scope documentation are provided to support that evidence during audit.

  • What is an error map, and can it be updated on my controller?

    The error map is the controller's volumetric compensation table, built from geometric measurement across the machine's volume. It can be updated during geometric adjustment on most controllers; some OEM-locked controllers restrict this and require factory-level access.

Further Detail

Standards, traceability and equipment control are set out on the quality system page. Commercial terms are in the terms and conditions.