Detroit CMM Calibration and Repair

Primary Service

CMM Calibration and Repair

Request a quote for verification or repair.

  • Work performed under ISO/IEC 17025:2017
  • Serving Michigan and the Greater Midwest
  • As-found data recorded before adjustment
  • Certificate states method, conditions and uncertainty

Request a quotation

A written quotation follows by email.

Verification, geometric adjustment, repair

The X, Y and Z scales are inspected and cleaned first, followed by a reader head check and an air bearing and airline inspection. Geometry — straightness, linearity and squareness in the XY, ZX and ZY planes — is measured next.

Length measurement error is then tested per ISO 10360-2:2009: five calibrated lengths, three repeats, seven positions, 105 measurements against E0,MPE. Probing error follows ISO 10360-5, PFTU, 25 points on a dedicated test sphere separate from the probe's reference sphere.

What a reverification visit covers

Stamping-floor CMMs see coolant, chip load and frequent probe changes, so mechanical checks run before the error tests. Where a fourth axis is fitted, rotary table performance is verified separately under ISO 10360-3:2000, using two test spheres at a specified height difference and radius.

  1. Scale and reader head

    Glass or steel scales and reader heads inspected and adjusted to correct output.

  2. Air bearings and drives

    Airline pressure, filtration and bearing condition checked before geometry is measured.

  3. Length measurement error

    E0 and E150 tested against gauge blocks, step gauges or a laser interferometer.

  4. Probing error (PFTU)

    Single-stylus form error measured on a calibrated test sphere per ISO 10360-5.

  5. Volumetric compensation

    Error-map update where the controller allows access; some OEM-locked maps require factory service.

Sequence of a calibration visit

  1. Thermal soak

    Machine and artifacts given time at ambient before any test begins.

  2. Mechanical check

    Scales, reader heads, air bearings and airlines inspected and cleaned.

  3. As-found testing

    Geometry, E0/E150 and probing error measured before adjustment.

  4. Adjustment and as-left

    Geometric correction and compensation update, then as-left verification and certificate issue.

Why Detroit CMM Calibration and Repair

  1. Repair and verification together

    Probe, drive, scale and air bearing repair are arranged alongside the same visit as verification.

  2. OEM-locked controllers addressed

    Where a controller restricts error-map access, that limitation is identified before the quote is finalized.

  3. Crash recovery and relocation

    Machines moved or crashed are re-verified against the same standard cited on the prior certificate.

Coverage across Michigan and the Greater Midwest

Service covers Detroit and surrounding Michigan and Greater Midwest markets. On-site availability varies by distance and schedule; confirm coverage for a specific address through a quote request.